The system operates similar to the FAPE/+60, but is equipped with two test vessels, which can be used independently from each other. The interface displays the testing conditions of both vessels (one indication for each test vessel).
  • The test space air is heated to a constant temperature of +60 °C ±0.5 K and purged continuously with 60 l/h of temperature pre-conditioned air.
  • The test chamber is designed as a gas tight area. It is temperature conditioned via the walls and a water pump circuit.
  • Leak-proofness of the test chamber is monitored via a pressure sensor and displayed on the touch-panel (operating console).
  • Setting of the overpressure is effected manually by means of a manual control valve (1000 – 1200 Pa).
  • In the test chamber, formaldehyde gas is mixed with the inlet air and led into washing bottles filled with demineralized water via the air outlet. The formaldehyde which has gassed out is dissolved in the water.
  • The routing of the outgoing air is effected by a five-fold distribution with solenoid valves. The air can be released freely into the atmosphere or it is routed via four separate pairs of gas washing bottles (not included in the scope of delivery) which are controlled by a timer.