Low Temperature Chamber – LTC – LN2 – 16m<sup>3</sup> – (565ft<sup>3</sup>)
Low Temperature Chamber – LTC – LN2 – 16m<sup>3</sup> – (565ft<sup>3</sup>)

Low Temperature Chamber – LTC – LN2 – 16m3 – (565ft3)

Climatic chamber 16m3 (565ft3) Hot Cold

-180°C / +200°C

 
Industries
Space and electronics  
Application
Thermal tests on TX and RX antennas
Description
Unit composed of :
  • The test chamber
  • The machinery on the rear with its electrical cabinet
  • Rear removable partial wall and specific feedthrough, for RF signals (1500x1500mm)
  • Control and supervision by SPIRALE Vs software
 
Technical Data
Performances are given for :
  • An ambient temperature between +15°C and +35°C
  • A load of 150 kg
  • A dissipation of 500 W
  Temperature
  • Temperature range : -180°C to +200°C
  • Cooling rate : average 1°C/min, with the load
  • Heating rate : average 1°C/min, with the load
  • Temperature stability : ≤ ± 0,5°C
  • Temperature homogeneity : ± 1,5°C at 20 cm of the walls
(According to NFX15 140 / IEC 60068-3-5)   Test  space  dimensions : WxDxH : 2500mm x 2500mm x 2500mm (98.4” x 98.4” x 98.4”)