TAH – Temperature Altitude Humidity Chamber 1m<sup>3</sup> (35ft<sup>3</sup>)
TAH – Temperature Altitude Humidity Chamber 1m<sup>3</sup> (35ft<sup>3</sup>)

TAH – Temperature Altitude Humidity Chamber 1m3 (35ft3)

Climatic chamber 1m3 (35ft3) Hot Cold Vacuum Humidity

-70°C / +180°C

1 mbar (approx. altitude 106.4 kft)

 

Industries

Aeronautics, electronics and defense  

Application

Climatic and altitude tests on electronic equipment for aeronautic industry  

Description of the equipment

  • Test chamber with its conditioning unit at the back
  • Machinery on the rear side with its electrical cabinet
  • Command panel on the side
  • Control and supervision software SPIRALE Vs

Technical Data

Performances are given for :
  • An ambient temperature between +15°C and +35°C
  • Water temperature inlet / outlet : +20°C / +25  °C
  • A constituted load of : 0kg
  • A thermal load of : 0 kW
  Temperature
  • From atmospheric pressure to 200 mbar (approx. altitude 38.5 kft)
  • Range : -70°C /   +180°C
  • Cooling rate 5°C/mn average from +85°C to –55°C 3°C/mn average from +100°C to –60°C
  • Heating rate 5°C/mn average from –55°C to +85°C 3°C/mn average from –60°C to +100°C At atmospheric pressure
  • Stability : ≤  ±0,5°C
  • Homogeneity : ± 1°C to 3°C at 10cm of the walls Control accuracy : ±0,2°C
Humidity
  • At atmospheric pressure
  • Humidity range : 10% to 98% without dissipation
  • For temperature range : +10°C /   +85°C
  • Humidity stability : ≤ ± 5 % RH
  • Control accuracy : ± 1% RH
  • Dew point range : +5 .. +  79°C
Vacuum
  • Range : from AP to 1 mbar, regulated down to 20 mbar
  • Variation speed : from AP to 20 mbar in 30 min
  • Control accuracy : ± 2 mbar from AP to 10 mbar
Climatic chamber Dimensions TAH - Temperature Altitude Humidity Chamber