TAH – Temperature Altitude Humidity Chamber 500L (18ft<sup>3</sup>)
TAH – Temperature Altitude Humidity Chamber 500L (18ft<sup>3</sup>)

TAH – Temperature Altitude Humidity Chamber 500L (18ft3)

Climatic chamber 500L (18ft3) Heat Cold Vacuum

Humidity -70°C / +100°C

10 mbar (approx. altitude 85 kft)

     

Industries

Aeronautics, electronics and defense  

Application

Climatic and altitude tests on electronic equipment for aeronautic industry  

Description of the equipment

Unit  composed  of :
  • The test chamber with its conditioning unit at the back
  • Machinery on the rear side with its electrical cabinet
  • Command panel on the side
  • Control and supervision software SPIRALE Vs

Technical Data

Performances are given for :
  • An ambient temperature between +15°C and +35°C
  • Water temperature inlet / outlet : +20°C / +25  °C
  • A constituted load of : 0kg
  • A thermal load of : 0 kW
  Temperature
  • From atmospheric pressure to 200 mbar (approx. altitude 38.5 kft)
  • Range : From -70°C to +100°C
  • Cooling rate 5°C/mn average from +100°C to –70°C
  • Heating rate 5°C/mn average from –70°C to +100°C At atmospheric pressure
  • Stability : ≤  ±1°C
  • Homogeneity : ± 1°C to 3°C at 25cm of the walls
  Humidity
  • At atmospheric pressure (site level)
  • Humidity range : 20% to 95% without dissipation For temperature range : +20°C /   +80°C
  • Humidity stability : ≤ ± 5 % RH Control accuracy : ± 1% RH
  Vacuum
  • Range : from AP to 10 mbar, controlled down  to  50  mbar  (approx.  altitude  63.5  kft)
  • Variation speed : from AP to 20 mbar in 30 mn
  • Control accuracy : ± 3 mbar from AP to  10mbar
Dimensions Climatic chamber 500L (18ft3) Heat Cold Vacuum