TVC – Space Simulator 700L
TVC – Space Simulator 700L

TVC – Space Simulator 700L

Vacuum chamber 700L 10-6 mbar

-50°C / +100°C

   

Industries

Space and electronics  

Application

Perform thermal cycling tests under secondary vacuum conditions on electronic components for space applications  

Description of the equipment

  • A cylindrical vacuum chamber with thermal plates and shrouds independently regulated in temperature
  • A frame in which the machinery and pumps are installed
  • A control panel on the electrical cabinet
  • A supervision and control system SPIRALE Vs

Technical Data

Performances are given for :
  • An ambient temperature between +15°C and 25°C
  • A dissipation : 200 W per thermal plate
  Temperature
  • Range : -50°C to +100°C
  • Average cooling rate : 3°C/mn without load
  • Average heating rate : 3°C/mn without load
  • Temperature stability : ± 1°C
  • Homogeneity on the shrouds : ± 5°C
  • Homogeneity on thermal plate : ± 1,5°C
  Vacuum
  • Range : from AP to 10-6 mbar
  • Variation speed : from AP to 2.10-5 mbar in 2h
  Space Simulator Vacuum chamber 700L Dimensions Space Simulator Vacuum chamber 700L