TS-60 Thermal Shock Test Chamber

The TS-60 small thermal shock test chamber is ideal for the testing of IC chips and small electronics. Design is constructed of two independently controlled chambers. One chamber is hot and the other one is cold. Product is rapidly transferred between the hot and cold chamber creating a temperature shock of the specimens. The TS Series thermal shock test chamber is the right solution for all your thermal shock testing requirements. These thermal shock chambers meet a variety of test specifications including Mil-Std 810, Mil-Std 883, IEC 60068-2-14, JEDEC and a variety of commercial test standards.

These TS Series Thermal Shock chambers are also in stock! click here.

  • Small Footprint
  • Energy-saving design
  • Quiet Operation at 58 dB(A)
  • More than 1,000 cycles without defrosting
  • The innovative WEBSeason® user interface allows you to program, control and monitor your tests at anytime and anywhere – even from your tablet or smartphone

Specifications

 

Model TS 60
Test Space Volume
(Cu.Ft/Liters)
2.1/60
Heating Rate  Up to 3.2°C/min
Cooling Rate  Up to 3.7°C/min
Temperature Range Hot Chamber: +50°C to +220°C

Cold Chamber: -80°C to +70°C

External Width 35.5″ (900mm)
External Depth 87″ (2150mm)
External Height 92″ (2330mm)

 

Product Features

Environmental conditions have a considerable effect on the function and the reliability of electronic components, devices and systems. The main purpose of temperature shock tests is to determine, if sudden temperature changes have an influence on the specimen’s long term function. In addition, early failures due to part and workmanship defects can be provoked without reaching the mechanical and thermal stress limits of the construction. 

  • Test space 2.1 cu. ft/60 liters
  • Vertical order of the chamber for safe and quick test performance
  • Regular temperature distribution in test space via a high recirculating rate with optimum air flow
  • Hot chamber temperatures of +50 °C to +220 °C
  • Cold chamber temperatures of –80 °C to +70 °C
  • Alternating time between the chambers <10 sec.
  • More than 1,000 cycles are possible without defrosting via a standard integrated volume compensation system for continuously operation
  • Low sound-pressure level, only 58 dB(A)
  • Operator friendly design enhances maintenance of the machine components
  • Entry port Ø 80 mm
  • Loading capacity of the lift 20 kg
  • 8″ TFT-color touch screen display
  • Control panel
  • Ready to plug-in
  • Window in hot zone
  • Ultra-lightweight shelf includes rails
  • Lift in loading position locked
  • Entry port Ø 80 mm
  • Volume compensation system for long-term operation
  • Air-cooled refrigeration unit
  • Hermetically sealed CFC-free refrigeration circuits
  • 8″ TFT-color touch screen display
  • CONTROLPAD* for indication of actual values
  • 32 bit control and monitoring system S!MPAC*
  • Software S!MCONTROL* for comfortable operation
  • Interfaces RS 232/USB/Ethernet
  • Digital I/O, potential-free, 24 V, 4 freely available inputs/outputs
  • Independent adjustable temperature limiter (tmin/tmax ) for hot and cold zone
  • Potential-free contact for switching-off of test specimens
  • Adjustable software temperature limiter min/max
  • Hours of operation counter.
  • Temperature control via sensor in cradle or in hot or cold zone
  • Defrosting cycles: automatic and programmable
  • Delayed start option
  • Standards are pre-programmed and recallable
  • Mobile version
  • Calibration of two temperature values
  • Powder-coated housing made of galvanized steel sheet
  • Test space, lift and insert shelves (insert baskets) made of corrosion resistant stainless steel

Technical Data

Type TS 60 TS 120 TS 120 P TS 300 TS 300 P
External Dimensions WxDxH | mm/inches 895x1970x1895

35.2×77.5×74.6

1990x2350x1985

39×92.5×78.2

990x2625x1985

39×103.5×78.2

2269x2377x2230

89.5×93.5×88

2469x2775x223

97×109.5×88

Test basket Volume | liters/cu ft 60/2.1 120/4.2 120/4.2 300/10.5 300/10.5
Test Basket Dimensions WxDxH | mm/inches 380x430x370

15x17x14.5

470x650x410

18.5×25.6×16.1

470x650x410

18.5×25.6×16.1

770x650x610

30.3×25.5×24

770x650x610

30.3×25.5×24

Temperature range

hot chamber | °C/°F

+50 to +220

+122 to +428

+50 to +220

+122 to +428

+50 to +220

+122 to +428

+50 to +220

+122 to +428

+50 to +220

+122 to +428

Temperature range

cold chamber | °C/°F

-80 to +70

-112 to +158

-80 to +70

-112 to +158

-80 to +70

-112 to +158

-80 to +70

-112 to +158

-80 to +70

-112 to +158

Heating Rate

hot chamber | °C/min

17.0 14.0 18.0 11.0 12.0
Cooling Rate

cold chamber | °C/min

3.7 6.3 7.5 5.0 5.0
Heating rate of cold chamber | °C/Min 3.2 2.0 2.0 1.5 1.5
Temperature Deviation, in time | °C ±0.3 to ±1.0 ±0.3 to ±1.0 ±0.3 to ±1.0 ±0.3 to ±1.0 ±1.0
Basket transfer time | sec <10 <10 <10 <10 <10
Recovery time | Min <15 <15 <12 <15 <15
Refrigeration Air-Cooled Water-Cooled

Additional Options

  • Analogue transducer l/O
  • Temp. measuring on test specimen
  • Temp. range extension to +250 °C
  • Wire mesh and insert shelves
  • Additional entry port Ø 80 mm
  • Connection for GN2
  • Shock cooling with LN2
  • Special voltages
  • WKD or DKD calibrations
  • Interface RS 232 <–> lEEE 488
  • Interface RS 422/485 (network card for test cabinet)
  • Software Simpati* for network operation

Brochures

Videos

TS  Series Thermal Shock test chamber from Weiss Technik that provides performance and reliability for all your testing requirements.
Click on link below for TS Series video:

Product Inquiry

Phone

1-800-368-4768

Address

3881 N. Greenbrooke Dr. SE